Fluke Calibration
Fluke 5502A Multi-Product Calibrator
Model No :
Fluke 5502A Multi-Product Calibrator
Part Code :
3843231

Overview

The 5502A Multi-Product Calibrator replaces the 5500A and addresses common workload items like 3.5 and 4.5 digit digital multimeters and more.
The 5502A sources direct voltage and current, alternating voltage and current with multiple waveforms and harmonics, simultaneous voltage outputs or dual voltage and current to simulate dc and ac power with phase control, resistance, capacitance, thermocouples and RTDs. The 5502A can also measure thermocouples and thermocouple simulators.

Features
- A full-functioned calibrator that covers a wide range of common workload
- Priority software support helps you stay productive
- Calibration and repair service
- Automate with MET/CAL Plus software for consistent and efficient calibration
- Robust protection circuits prevent costly damage from operator error
- Ergonomically designed carrying handles make the 5502A easy to transport
- Current output that extends to 120 A when paired with the 52120A Transconductance Amplifier

Major Applications
- Calibration of following instruments:
- Handheld and bench meters (analog and digital) to 4.5 digits
- Current clamps and clamp meters
- Panel meters
- Electronic thermometers
- Chart recorders
- Oscilloscope recorders
- XY recorders
- Data loggers

Technical Specifications
- Function & range:
• Direct volts: 0 to ± 1020 V
• Direct current: 0 to 20.5 A
• Alternating volts: 1 mV to 1020 V , 10 Hz to 500 kHz
• Volt*Hertz: 1000 V@ 10 kHz/330 V@100 kHz
• Alternating current: 29 µA to 20.5 A ,10 Hz to 30 kHz
• Waveforms: Sine, square, triangle, truncated sine
• Resistance: 0 O to 1100 MO
• Capacitance: 220 pF to 110 mF
• Thermocouple (source and measure temperature): B, C, E, J, K L N R, S, T, U 10 µV/°C and 1 mV/degC
• RTD (source temperature): Pt 385-100 O, Pt 3926-100 O Pt 3916-100 O,
Pt 385-200 O, Pt 385-500 O, Pt385 1000 O, PtNi 385-120 O, (Ni120), Cu 427 10 O

General Specifications
- Interfaces: RS-232, IEEE 488
- Frequency uncertainty: <25 ppm
- Calibration Documentation: 17025 accredited report of calibration included
- Oscilloscope calibrator (optional): Levelled sine wave from 5 mV to 5.5 Vpp max, frequencies 50 kHz to 600 kHz; edge rise times of < 300 ps, multiple trigger functions, lowest dc, square wave and timing uncertainty
- Current boost transconductance amplifier: 0 to 100 A dc, 0 to 120 A ac, dc to 10 kHz, up to 6000 A with accessory current coils